Fluke FETI125/C3I-30HZ - CNXTM Ti125 with Three i00 iFlex Clamp Modules
Capture three True RMS current measurements in hard to reach places on your next infrared inspection with the Fluke CNX Ti125 iFlex® kit. The i3000 iFlex® modules allow you to measure around tight, awkward spaces and/ or sizes and makes wire access easier. You can then view your current measurement results on the Ti125 thermal imaging camera. Use the modules as stand-alone measurement tools or combine to capture up to three critical measurements along with the infrared image. Simplify and streamline your reporting with our easy to use SmartView® Software. Download the SmartView Mobile app and you can use your Apple® iPhone® or iPad® to generate a report and communicate your results directly to your client or colleague from the jobsite. Work smarter and faster!
Get all the information you need to quickly troubleshoot problems with the Fluke CNX Ti125 Industrial Kit.
The Ti125 Offers:
- IR-Fusion® with AutoBlend- blends a digital and partially transparent IR image into a single image, making problem detection fast and easy.
- Exclusive IR-OptiFlex™ focus system—ensures that images are in good focus from 1.2 meters (4 feet) and beyond for optimum image clarity and scanning convenience. For shorter distances change to manual mode with the touch of a finger.
- IR-PhotoNotes™ annotation system—Quickly identify and keep track of inspection locations by adding digital images of important information and surrounding areas.
- Multi-mode video recording—focus-free video in visible light and infrared with full IR-Fusion.
- Easily communicate the location of problems with the Electronic Compass (8-Point Cardinal).
The CNX i3000 iFlex® AC Wireless Current Module Offers:
- True-rms AC current measurements to 2500 A
- 3% accuracy
- Memory recording up to 65,000 readings
- iFlex current probe allows you to get into tight, awkward spaces
- Backlit LCD display
- CAT III 1000, CAT IV 600 V
- IP42
Fluke Ti125 Infrared Camera
| Detailed Specifications | |
| IR resolution (FPA size) | 160 x 120 FPA Uncooled Microbolometer |
| Spectral band | 7.5 μm to 14 μm (long wave) |
| Capture or refresh rate | 9 Hz or 30 Hz versions |
| NETD (thermal sensitivity) | ≤ 0.10 °C at 30 °C target temp (100 mK) 0.10 °C at 30 °C target temp (100 mK) |
| FOV (field of view) | 22.5 °H x 31 °V |
| IFOV (spatial resolution) | 3.39 mRad |
| Temperature measurement range (not calibrated below -10 °C) | -20 °C to +350 °C (-4 °F to +662 °F) |
| Temperature measurement accuracy | ± 2 °C or 2 % (at 25 °C nominal, whichever is greater) |
| Focus mechanism | IR-OptiFlex™ focus system |
| IR-Fusion® technology | PIP, FULL IR, FULL VISIBLE, AutoBlend™ |
| CNX wireless enabled | Yes |
| Color alarms | High temperature, low temperature, isotherm |
| Standard palettes | Blue-red, grayscale, inverted grayscale, high-contrast, hot metal, ironbow, amber, inverted amber |
| Ultra Contrast™ palettes | Blue-red, grayscale, inverted grayscale, high-contrast, hot metal, ironbow, amber, inverted amber |
| Hot/cold markers | Yes |
| User definable spot markers | Three on camera and in SmartView® |
| Centerbox (MIN/AVG/MAX) | Yes |
| Level and span control | Manual and auto |
| Minimum span in auto mode | 5 °C |
| Minimum span in manual mode | 2.5 °C |
| Minimum IR focus distance | 15.25 cm (6 in) |
| Weight | 0.726 kg (1.6 lb) |
| Size | 28.4 x 8.6 x 13.5 cm (11.2 x 3.4 x 5.3 in) |
| LCD display | 3.5 inch diagonal (portrait format) |
| Visible camera | 2 megapixel industrial-grade |
| Minimum automatic parallax correction | ~18-22 inches |
| IR-PhotoNotes™ annotation system | Yes (3 images) |
| Laser pointer | Yes |
| Torch | Yes |
| Electronic (cardinal) compass | Yes |
| Emissivity correction | Yes |
| Transmission correction | Yes |
| Background (reflected) compensation | Yes |
| Voice annotation (audio) | Yes (60 seconds) per image |
| Multi-mode video output | Streaming USB video output (infrared, visible and IR-Fusion modes) |
| Multi-mode video recording (standard avi w/ mpeg encoding) | Yes (AVI with MPEG encoding, up to 5 minutes) |
| Multi-mode video recording (radiometric .Is3) | Yes, radiometric .is3 for approx. 2.5 to 5 minutes depending upon thermal scene |
| Memory review | Thumbnail review |
| Battery (field-replaceable, rechargeable) | Two |
| Battery life | 4+ hours (each)* |
| External battery charging base | Yes |
| Charging power supply | Yes |
| Drop test | 2 meter (6.5 feet) |
| Ingress protection (IP) rating (IEC 60529) | IP 54 |
| Recommended calibration cycle | Two-years |
| Multifunction card reader | Included |
| Memory storage | 2 GB SD memory card |
| Direct download capability | Mini USB download direct to PC |
| Operating temperature range | -10 °C to +50 °C (14 °F to 122 °F) |
| Storage temperature range | -20 °C to +50 °C (-4 °F to 122 °F) |
| Operating humidity | Operating and storage 10 % to 95 %, non-condensing |
| Vibration and shock | 2G, IEC 68-2-26 and 25G, IEC 68-2-29 |
| Safety standards | CSA (US and CAN): C22.2 No. 61010-1-04, UL: UL STD 61010-1 (2nd Edition), ISA: 82.02.01 |
| C Tick | IEC/EN 61326-1 |
| EMI, RFI, EMC | EN61326-1; FCC Part 5 |
| User manuals | Czech, English, Finnish, French, German, Italian, Japanese, Korean, Polish, Portuguese, Russian, Simplified Chinese, Spanish, Swedish, Traditional Chinese, Turkish, Dutch, and Hungarian |
| Standard warranty period | Two-years |
| Extended warranty and service plans | Yes |
*Assumes 50% brightness of LCD
Fluke CNX i3000 iFlex AC Wireless Current Module
| Specifications | |
| Range | 2500 A AC |
| Resolution | 0.1A to 999.9, 1A from 1000 to 2500 |
| Accuracy | 3 % ± 5 digits |
| Crest factor (50 Hz/60 Hz) | 3.0 at 1100 A, 2.5 at 1400 A, 1.42 at 2500 A, add 2 % for C.F. > 2 |
| Display | 3-1/2 digits, LCD w/backlight |
| Log rate/interval | 1 sec minimum/adjustable by PC or front panel |
| Battery type | 2 AA, NEDA 15 A, IEC LR6 |
| Battery life | 400 hours |
| Memory record | up to 65,000 readings |
| Unbound listen interval | 5 seconds |
| RF communications | 2.4 GHZ ISM Band |
| RF communication range | 20 Meters |
| Operating temperature | -10 °C to +50 °C |
| Storage temperature | -40 °C to +60 °C |
| Operating humidity | 90 % at 35 °C, 75 % at 40 °C, 45 % at 50 °C |
| EMC | EN 61326-1:2006 |
| Safety compliance | EN/IEC 61010-1:2010 to 1000 V Measurement Category (CAT) III 600 V Measurement Category (CAT) IV EN/IEC 61010-2-030:2010 EN/IEC 610101-2-031:2002 EN/IEC 61010-2-032:2002 |
| Pollution degree | 2 |
| Temperature coefficient | Add 0.1 X (specified accuracy)/ °C (<18 °C or >28 °C) |
| Safety rating | CAT IV 600 V, CAT III 1000 V |
| Certifications | CSA, FCC T68-FWCS IC:6627A-FWCS |
| Ingress Protection (IP) rating | IP42 |
| Size (HxWxD) | 16.5 cm x 6.35 cm x 3.56 cm |
| (6.5 in x 2.5 in x 1.4 in) | |
| Weight | .22 kg (8 oz) |
| Jaw opening | 25.4 cm (10 in) coil |
Do you have a CNXTM Ti125 with Three i00 iFlex Clamp Modules that needs to be calibrated?
We offer standard traceable calibration and ISO 17025 accredited calibration services.
Get your CNXTM Ti125 with Three i00 iFlex Clamp Modules calibrated today! Click Here!