Your Cart
(empty)
0
Tradeport Electronics
measurement is our business
Cart : (empty)

GW Instek PCS-1000 - Precision Current Shunt Meter

Precision Current Shunt Meter

GW Instek

Rated 0 / 5 (0 reviews)

Category : DC Electronic Loads

by quote only add to cart

Documents:
Datasheet User's Manual

Overview
Features
Specifications
Calibration

PCS-1000 is a High-Precision D.C. and A.C. Current Shunt Meter which carries built-in current shunts and high-accuracy current measurement circuits. PCS-1000 not only measures current in five current levels ─300A, 30A, 3A, 300mA and 30mA, but also measures voltage at 6 1/2-digit resolution simultaneously. This unique feature puts PCS-1000 in an unprecedented position among all the shunt meters available in the market today. After a long time operation, PCS-1000 is able to maintain high measurement accuracy in spite of the temperature variation. This is due to its core design based on the high stability and low temperature coefficient shunts. PCS-1000, a verification-lab level current and voltage measurement equipment, is suitable for the R&D, quality verification and testing jobs of all power products such as power supply, communications system power, server power and electronic load. The maximum voltage/current measurement ranges of PCS-1000, including DC 1000V / AC 600V / AC 300A / DC 300A, meet all current and voltage measurement challenges of every power industry.

This product has been discontinued. Our replacement recommendation is the PCS-1000I - GW Instek Precision Current Shunt Meter

  • 6 1/2 Digit measurement resolution
  • Dual measurement for Voltage and Current
  • AC&DC Current measurement range:30mA/300mA/3A/30A/300A(Max)
  • AC Voltage measurement range:200mV/2V/20V/200V/600V(Max)
  • DC Voltage measurement range:200mV/2V/20V/200V/1000V(Max)
  • Auto zero circuit
  • Standard Interfaces: USB, GPIB
DC Voltage Characteristics
  Range 23℃±5℃(1 Year) Temperature Cofficient / ℃
  200.0000mV 0.0050+0.0035 0.0005+0.0005
  2.000000V 0.0050+0.0010 0.0005+0.0001
  20.00000V 0.0050+0.0010 0.0005+0.0001
  200.0000V 0.0050+0.0010 0.0005+0.0001
  1000.000V 0.0050+0.0020 0.0005+0.0001
Accuracy ±(% of reading +% of Range); Voltage input Resistance:10MΩ for all DC Voltage ranges
DC Current Characteristics
  Range Burden Voltage 23℃±5℃(1 Year) Temperature Cofficient / ℃
  30.00000mA <0.4V 0.01+0.005 0.001+0.002
  300.0000mA <0.5V 0.01+0.005 0.001+0.002
  3.000000A <0.8V 0.01+0.005 0.001+0.002
  30.00000A <0.8V 0.01+0.005 0.001+0.002
  300.0000A <0.8V 0.02+0.005 0.001+0.002
Accuracy ±(% of reading +% of Range) 
True RMS AC Voltage Characteristics
  Range Frequency 23℃±5℃(1 Year) Temperature Cofficient / ℃
  200.0000mV 45Hz-2KHz
2KHz-10kHz
10kHz-20kHz
0.5+0.05
1.0+0.05
2.0+0.10
0.005+0.005
  2.000000V 0.005+0.005
  20.00000V 0.005+0.005
  200.0000V 0.005+0.005
  600.000V 0.005+0.005
Accuracy ±(% of reading +% of Range) ℃
True RMS AC Current(AC+DC) Characteristics
  30.00000mA
45Hz-2KHz
2KHz-10kHz
0.5+0.05
1.0+0.05
0.03+0.006
  300.0000mA 0.03+0.006
  3.000000A 0.03+0.006
  30.00000A 45Hz-400Hz 0.5+0.05 0.03+0.006
  300.0000A 0.03+0.006
Accuracy ±(% of reading +% of Range) 
Current Monitor Accuracy
Range Shunt Value DC Accuracy AC Accuracy ≦ 400Hz Max Input DC/ AC RMS
  10Ω 0.01% 0.1% 30mA
  0.01% 0.1% 300mA
  0.1Ω 0.01% 0.1% 3A
  0.01Ω 0.01% 0.1% 30A
  0.001Ω 0.02% 0.1% 300A
General 
Power Supply 100V/120V/220V/240V±10%
Power Line Frequency 50/60Hz
Operating Environment Full Accuracy for 0℃ to 50℃, Full accuracy to 80% R.H. at 40℃      
Storage Environment -40℃ to 70℃
Power Consumprion Max 35VA
Dimensions 210 mm (W) x 80mm(H) x 390 mm(D)
Weight Approx 6kg

Do you have a Precision Current Shunt Meter that needs to be calibrated?

We offer standard traceable calibration and ISO 17025 accredited calibration services.

Get your Precision Current Shunt Meter calibrated today! Click Here!

 

There are currently no reviews for this product.

 

Review the GW Instek PCS-1000 - Precision Current Shunt Meter

Precision Current Shunt Meter

Name :

Email :

Title :

Review: