Your Cart
(empty)
0
Tradeport Electronics
measurement is our business
Cart : (empty)

Metrel MI 3152H - EUROTEST XC 2,5 KV

EUROTEST XC 2,5 KV

Metrel

Category : Electrical Testers

by quote only Add to Cart

Documents:
Users Manual Datasheet PC Software

Overview
Features
Information
Calibration

MI 3152H EurotestXC 2,5 kV is an instrument from the new generation of Metrel’s multifunctional measuring instruments. The already well known functions like complete installation safety testing according to IEC/EN 61557 and AUTO SEQUENCE testing of TN and TT earthing systems expanded with insulation resistance measurement with the test voltage up to 2,5 kV are managed by a completely new user interface based on large colour touch screen display. A wide range of functions is included: from on-line voltage monitoring, phase sequence testing, varistor testing, earth resistance measurement with 3 wire and 2 clamps method and illuminance measurement, TRMS current measurement up to RCD tests, line and loop impedance tests as well as diagnostic test enabled by PI and DAR indexes calculation.

  • Predefined profile dependent
  • AUTO SEQUENCEs.
  • Predefined Automatic tests:
  • Auto TT (U, Zln, Zs, Uc);
  • Auto TN/RCD (U, Zln, Zs, Rpe);
  • Auto TN (U, Zln, Zlpe, Rpe).
  • Built-in help screens for referencing on site.
  • Built-in fuse tables for automatic evaluation of the line / loop impedance result.
  • Monitoring of all 3 voltages in real-time.
  • Automatic polarity reversal on continuity test.
  • Automated RCD testing procedure.
  • Automated Impedance testing procedure (Z AUTO).
  • Functional inspections.
  • Measurement filtering according  to the selected area group.
  • Built-in charger and rechargeable batteries as standard accessory.
  • BT communication with PC, Android tablets and smart phones via built-in BT.
  • PC SW Metrel ES Manager (structure and report creation, data upload/download).
  • Optional aMESM Android APP (structure and report creation, data upload/download).

Continuity

Function Measuring Range Resolution Accuracy
Test Current 7 mA (2-wire) 0.00 Ω ... 19.99 Ω
20.0 Ω ... 1999 Ω
0.1 Ω
1 Ω
±(5 % of r. + 3 digits)
Test Current 200 mA (2-wire) 0.00 Ω ... 19.99 Ω
20.0 Ω ... 199.9 Ω
200.0 Ω ... 1999 Ω
0.01 Ω
0.1 Ω
1 Ω
±(3 % of r. + 3 digits)
±(5 % of r.)
±(5 % of r.)

Insulation Resistance

Function Measuring Range Resolution Accuracy
Test Voltage
50/100/250 V
0.00 MΩ ... 19.99 MΩ
20.0 MΩ ... 99.9 MΩ
100.0 MΩ ... 199.9 MΩ
0.01 MΩ

0.1 MΩ
±(5 % of r. + 3 digits)
±(10 % of r.)
±(20 % of r.)
Test Voltage
500/1000 V
0.00 MΩ ... 19.99 MΩ
20.0 MΩ ... 199.9 MΩ
200 MΩ ... 999 MΩ
0.01 MΩ
0.1 MΩ
1 MΩ
±(5 % of r. + 3 digits)
±(5 % of r.)
±(10 % of r.)
Test Voltage
2500 V
0.00 MΩ... 19.99 MΩ
20.0 MΩ... 199.9 MΩ
200 MΩ... 999 MΩ
1.00 GΩ... 19.99 GΩ
0.01 MΩ
0.1 MΩ
1 MΩ
0.01 GΩ
±(5 % of r. + 3 digits)
±(5 % of r.)
±(10 % of r.)
±(10 % of r.)

Insulation Analysing

Function Measuring Range Resolution Accuracy
Calculation of PI, DAR, DD Only for test voltage 500/1000/2500 V 0.01 MΩ... 9.99 MΩ
10.0 MΩ... 100 MΩ
0.01 MΩ
0.1 MΩ
±(5 % of r. + 2 digits)
±(5 % of r.)

RCD

Function Measuring Range Resolution Accuracy
RCD Uc 0.00 V ... 19.99 V
20.0 V ... 99.9 V
0.1 V -0%/+15 %) of r. ± 10 digits
(-0%/+15 %) of r.
RCD (t) 0.00 ms ... 40.0 ms
0.0 V ... max.time
0.1 ms ±1 ms
±3 ms
RCD I Ramp 0.2xI∆N ... 1.1xI∆N (AC)
0.2xI∆N ... 1.5xI∆N (A),
I∆N ≥30 mA)
0.2xI∆N ... 2.2xI∆N (A), I∆N <30 mA)
0.05xI∆N ±0.1xI∆N

Impedance

Function Measuring Range Resolution Accuracy
Zline L-L, L-N Ipsc 0.00 Ω ... 9.99 Ω
10.0 Ω ... 99.9 Ω
100 Ω ... 999 Ω
1.00 kΩ ... 9.99 kΩ
0.01 Ω
0.1 Ω
1 Ω
10 Ω
±(5 % of r. + 5 digits)

±(10 % of r.)
Zloop L-PE, Ipfc 0.00 Ω ... 9.99 Ω
10.0 Ω ... 99.9 Ω
100 Ω ... 999 Ω
1.00 kΩ ... 9.99 kΩ
0.01 Ω
0.1 Ω
1 Ω
10 Ω
±(5 % of r. + 5 digits)

±(10 % of r.)
 

Voltage

Function Measuring Range Resolution Accuracy
TRMS 0 … 550 V 1 V ±(2 % of r. + 2 digits)
Frequency 0.00 Hz .. 9.99 Hz
10.0 Hz ... 499.9 Hz
0.01 Hz
0.1 Hz
±(0.2 % of r. + 1 digits)

Current

Function Measuring Range Resolution Accuracy
TRMS, AC with A 1018 0.0 mA ... 99.9 mA
100 mA ... 999 mA
1.00 A ... 19.99 A
0.1 mA
1 mA
0.01 A
±(5 % of r. + 5 digits)
±(3 % of r. + 3 digits)
±(3 % of r.)
TRMS, AC with A 1019 0.0 mA ... 99.9 mA
100 mA ... 999 mA
1.00 A ... 19.99 A
0.1 mA
1 mA
0.01 A
Indicative
±(5 % of r.)
±(3 % of r.)
TRMS, AC/DC with A 1391, range=40 A 0.00 A ... 1.99 A
2.00 A ... 19.99 A
20.0 A ... 39.9 A
0.01 A
0.01 A
0.1 A
±(3 % of r. + 3 digits)
±(3 % of r.)
±(3 % of r.)
TRMS, AC/DC with A 1391, range = 300 A 0.00 A ... 19.99 A
20.0 A ... 39.9 A
40.0 A ... 299.9 A
0.01 A
0.1 A
0.1 A
Indicative
Indicative
±(3 % of r. + 5 digits)

Earth Resistance

Function Measuring Range Resolution Accuracy
3 wire 0.00 Ω ... 19.99 Ω
20.0 Ω ... 199.9 Ω
200.0 Ω ... 9999 Ω
0.01 Ω
0.1 Ω
1 Ω
±(5 % of r. + 5 digits)
2 clamp 0.00 Ω ... 19.99 Ω
20.0 Ω ... 30.0 Ω
30.1 Ω ... 39.9 Ω
0.01 Ω
0.1 Ω
0.1 Ω
±(10 % of r. + 10 digits)
±(20 % of r.)
±(30 % of r.)
Specific earth resistance 0.0 Ωm … 99.9 Ωm
100 Ωm ... 999 Ωm
1.00 kΩm ... 9.99 kΩm
10.0 kΩm ... 99.9 kΩm
100 kΩm ... 9999 kΩm
0.1 Ωm
1 Ωm
0.01 kΩm
0.1 kΩm
1 kΩm
±(5 % of r.) for Re 1 Ω … 1999kΩ
±(10 % of r.) for Re 2 kΩ … 19.99kΩ
±(20 % of r.) for Re > 20 kΩ

Illuminance

Function Measuring Range Resolution Accuracy
Type B 0.01 lux ... 19.99 lux
20.0 lux ... 199.9 lux
200 lux ... 1999 lux
2.00 klux ... 19.99 klux
0.01 lux
0.1 lux
1 lux
10 lux
±(5 % of r. + 2 digits)

±(5 % of r.)
 
Type C 0.01 lux ... 19.99 lux
20.0 lux ... 199.9 lux
200 lux ... 1999 lux
2.00 klux ... 19.99 klux
0.01 lux
0.1 lux
1 lux
10 lux
±(10 % of r. + 3 digits)

±(10 % of r.)
 
Specific earth resistance 0.0 Ωm … 99.9 Ωm
100 Ωm ... 999 Ωm
1.00 kΩm ... 9.99 kΩm
10.0 kΩm ... 99.9 kΩm
100 kΩm ... 9999 kΩm
0.1 Ωm
1 Ωm
0.01 kΩm
0.1 kΩm
1 kΩm
±(5 % of r.) for Re 1 Ω … 1999kΩ
±(10 % of r.) for Re 2 kΩ … 19.99kΩ
±(20 % of r.) for Re > 20 kΩ

General

Power Supply 9 VDC (6x1.5 V battery or accu, size AA)
Overvoltage category 1000 V DC CAT II; 600 V CAT III; 300 V CAT IV
Protection class double insulation
COM port BT, USB, RS232
Weight 1.3 kg
Size ( l x h x w ) 230 x 103 x 115 mm

Do you have a EUROTEST XC 2,5 KV that needs to be calibrated?

We offer standard traceable calibration and ISO 17025 accredited calibration services.

Get your EUROTEST XC 2,5 KV calibrated today! Click Here!