Soken DAC-PBC-8 - Electronic Burden for Current Transformer

Electronic Burden for Current Transformer

Soken

Category : Instrument Transformer Testing Set

by quote only Add to Cart

Documents:
Brochure

Overview
Features
Information
Calibration
  • Programmable burden used for current error testing in accordance with international standard IEC 60044-1.
  • Testing current can be varied over a wide range: from 1 % to 200% of the rated secondary current.
  • Rated burdens from 0 to 50VA are programmable.
  • USB Interface is a standard fixture.
  • Lead wire compensation (the maximum of 0.2 Ω) by connecting with four terminals is possible.
  • Max real burden is up to 400VA with option units. (when 200% at rated secondary current 1A)

 

4-terminal Connecting diagrams

 

Option Electric Burden

Optional burden DAC-FBC series are available in need of further VA.

  • DAC-FBC Series: : 50VA, 100VA and 200VA type

Combining DAC-PBC-8 with all FBC series, max 400VA* is available. *(when 200% at rated secondary current 1A)

 

Instrument Transformer Test Set DAC-VCTT-8

Combining with DAC-VCTT-8, Current Burden DAC-PBC-8 can be controlled from DAC-VCTT-8. Click here to find out more about the DAC-VCTT-8.

  • Programmable burden used for current error testing in accordance with international standard IEC 60044-1.
  • Testing current can be varied over a wide range: from 1 % to 200% of the rated secondary current.
  • Rated burdens from 0 to 50VA are programmable.
  • USB Interface is a standard fixture.
  • Lead wire compensation (the maximum of 0.2 Ω) by connecting with four terminals is possible.
  • Max real burden is up to 400VA with option units. (when 200% at rated secondary current 1A)

 

General

Dimensions W430 x H250 x D380mm (without projections)
Rated Supply Voltage100 to 240V ±10%
Other Functions Test condition programming, over current protection, Alarm function for measuring circuit, lead wire compensation
CommunicationUSB (2.0/1.1)
Accuracy (target value)Burden and Burden Power Factor: ±4%
Settable resolution
Burden Power Factor0.01 steps
Rated burden setting0 – 4 VA: 0.001 VA steps
4 – 20 VA: 0.01 VA steps
20 – 50 VA: 0.1 VA steps
Test Frequency50/60 Hz
Lead wire compensation0.000 to 0.200Ω
Burden Power factorDelay 0.5 to 1
Rated burden setting 0.000 to 50.0 VA
Maximum of 400VA* with optional Electric burden *1% to 200% of the rated secondary current
Test Current range1% to 200% of the rated secondary current
Rated Secondary Current1A and 5A